Phthalates
The method is primarily used in collaborative research projects, and other analytical services are agreed on a case by case basis. Written requests and contacts:
Environmental pollutant analysis
Accreditation
- The method is accredited for urine
Number of samples
- Minimum 6 samples
Matrices
- Urine
Method description
- Samples are pretreated enzymatically. The quantification is performed with a liquid chromatograph coupled to a triple quadrupole mass spectrometer (LC-MS/MS).
Limit of quantification
- 0.25–2.5 ng/ml
Sampling instructions
- Sample container: glass or polypropylene (PP)
- Minimum sample amount: 1.0 ml
- Maximum storage time: +7°C for 3 days, -18°C for at least 20 years, -70°C or in liquid nitrogen indefinitely
- See also the general sampling instructions that apply to all analyses:
Environmental pollutant analysis
Dispatch
- Frozen in cold package
Analytes
- Monoethylphthalate
- Monoisobutylphthalate
- Mono-n-butylphthalate
- Mono-(2-ethylhexyl)-phthalate
- Monobenzylphthalate
- Mono-(2-ethyl-5-oxohexyl)-phthalate
- Mono-(2-ethyl-5-hydroxyhexyl)-phthalate
- Mono-(2-ethyl-5-carboxypentyl)-phthalate
- Monocyclohexylphthalate
- Mono-n-octylphthalate
- Monoisononylphthalate
- Monocarboxy-isooctylphthalate
- Monohydroxy-isononylphthalate
- Mono(propyl-6-hydroxyheptyl)phthalate
- Monocarboxy-isononylphthalate
The unit is a FINAS accredited testing laboratory T077. It has been accredited according to the standard SFS-EN ISO/IEC 17025:2005 and its scope of accreditation is available on the website of FINAS (Finnish Accreditation Service).
The work order must be included in the shipment of samples along with an itemized sample list, if needed.
Work order for environmental pollutant analyses (PDF 132 KB)